자료실

메뉴하단배경

종합자료실library

[FlowCAM] 입자 이미지 분석 : 분해능 및 샘플링 고찰
작성자 : 삼보과학   작성일 : 2011-07-11   조회수 : 4329
파일첨부 : ResolutionWhitePaper.pdf
Imaging Particle Analysis: Resolution and Sampling Considerations
=============================================================
Abstract
Imaging particle analysis represents an exciting method of particle analysis which combines the speed of automated particle analyzers with the discrimination found in optical microscopy. However, in order to discriminate shape differences, it is limited to particles of a certain size or larger. This paper discusses the primary factors that lead to this limitation, caused by the optical system and the sensor. In the end, diffraction limits of the optical system and sampling limits incurred at the sensor limit this technique to particle counting of particles 1μm in Equivalent Spherical Diameter (ESD) and larger, and particle characterization of particles 2μm in ESD and larger. Due to this limitation, imaging particle analysis of submicron particles is limited to non-optical techniques such as electron microscopy.

* 상기 내용은 논문 'Imaging Particle Analysis: Resolution and Sampling Considerations'의 초록부분입니다. 전문을 보시기를 원하시는 분은 첨부화일을 다운로드하여 보시기 바랍니다.
이전글 [FlowCAM] 입자성 물질 측정용 디지털이미지분석기(Size & Shape)
다음글 [FlowCAM] 비경구 약배합과정에서의 단백질 응집 검출
       

아이콘    고객지원  >  종합자료실